Remote Control Commands
R&S
®
SMBV100A
607Operating Manual 1176.8016.02 ─ 17
Manual operation: See "Current Range without Recalculation" on page 401
[:SOURce<hw>]:BB:ARBitrary:TRIGger:RMODe?
The command queries the status of waveform output or all trigger modes with ARB on.
Return values:
<RMode> STOP | RUN
RUN
The waveform is output. A trigger event occurred in the triggered
mode.
STOP
The waveform is not output. A trigger event did not occur in the
triggered modes, or waveform output was stopped by the com-
mand:BB:ARB:TRIG:ARM:EXECute (armed trigger modes
only).
Example:
BB:ARB:TRIG:SOUR EXT
sets external triggering
BB:ARB:TRIG:MODE ARET
selects the Armed_Retrigger mode.
BB:ARB:TRIG:RMOD?
queries the current status of waveform output.
Response: RUN
the waveform is output, an external trigger was executed.
Usage: Query only
Manual operation: See "Running/Stopped" on page 391
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLENgth <SLength>
The command defines the length of the signal sequence to be output in the Single trig-
ger mode. The unit is defined with command SOUR:BB:ARB:TRIG:SLUNit. It is pos-
sible to output deliberately just part of the waveform, an exact sequence of the wave-
form, or a defined number of repetitions of the waveform.
Parameters:
<SLength> integer
Range: 1 to dynamic
*RST: 1
Default unit: sample
Example:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SAMP
sets unit Samples for the entry of sequence length.
BB:ARB:TRIG:SLEN 200
sets a sequence length of 200 samples. The first 200 samples of
the current waveform will be output after the next trigger event.
Manual operation: See "Signal Duration " on page 391