Remote Control Commands
R&S
®
SMBV100A
608Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLUNit <SLUnit>
The command defines the unit for the entry of the length of the signal sequence
(SOUR:BB:ARB:TRIG:SLEN) to be output in the "Single" trigger mode
(SOUR:BB:ARB:SEQ SING).
Parameters:
<SLUnit> SEQuence | SAMPle
*RST: SEQuence
Example:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SEQ
sets unit Sequence length for the entry of sequence length.
BB:ARB:TRIG:SLEN 2
sets a sequence length of 2 waveforms. The current waveforms
will be output twice after the next trigger event.
Manual operation: See "Signal Duration Unit" on page 390
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe <SMode>
The command selects the extended trigger mode for multi segment waveforms.
Parameters:
<SMode> SAME | NEXT | NSEam | SEQuencer
NSEam = Next Segment Seamless
*RST: NEXT
Example:
SOUR:BB:ARB:SEQ AUTO
selects trigger mode AUTO.
SOUR:BB:ARB:TRIG:SMOD SAME
the same segment of the waveform is output repeatedly.
Manual operation: See "Next Segment Mode" on page 394
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SOURce <Source>
Selects the trigger source.
Parameters:
<Source> INTernal|EXTernal
INTernal
manual trigger or *TRG.
EXTernal
trigger signal on the TRIG connector.
*RST: INTernal
Example:
SOURce1:BB:ARBitrary:TRIGger:SOURce EXTernal
sets external triggering via the TRIGGER connector.
Manual operation: See "Trigger Source" on page 391
SOURce Subsystem