Remote Control Commands
R&S
®
SMBV100A
612Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode> UNCHanged | RESTart | PULSe | PATTern | RATio | TRIGger
UNCHanged
A marker signal as defined in the waveform file (tag 'marker
mode x') is generated.
RESTart
A marker signal is generated at every waveform start.
PULSe
A pulsed marker signal is generated. The pulse frequency (=
sample rate/divider) is defined with the
SOUR:BB:ARB:TRIG:OUTP:PULS:DIV command and can be
queried with the SOUR:BB:ARB:TRIG:OUTP:PULS:FREQ?
command.
PATTern
A marker signal is generated with the aid of a user-definable bit
pattern. The bit pattern is entered with the aid of com-
mand :BB:ARB:TRIGger:OUTPut:PATTern. The bit pattern
is a maximum of 32 bits long.
RATio
A regular marker signal corresponding to the Time Off / Time On
specifications in the commands :ARB:TRIGger:OUTPut:
OFFTime and :ARB:TRIGger:OUTPut:ONTime is generated.
TRIGger
A received internal or external trigger signal is output at the
marker connector.
*RST:
UNCHanged
Example:
BB:ARB:TRIG:OUTP2:MODE PULS
selects the pulsed marker signal on output MARKER 2.
BB:ARB:TRIG:OUTP2:STAT ON
activates the pulsed marker signal on output MARKER 2.
Manual operation: See "Marker Mode" on page 401
[:SOURce<hw>]:BB:ARBitrary:TRIGger:OUTPut<ch>:OFFTime <OffTime>
[:SOURce<hw>]:BB:ARBitrary:TRIGger:OUTPut<ch>:ONTime <OnTime>
The command sets the number of samples in a period (ON time + OFF time) during
which the marker signal in setting :ARB:TRIGger:OUTPut:MODE RATio on the
marker outputs is ON.
Parameters:
<OnTime> integer
Range: 1 to 14913079
*RST: 1
Example:
BB:ARB:TRIG:OUTP2:ONT 20
sets an ON time of 20 samples for marker 2
Manual operation: See "Marker Mode" on page 401
SOURce Subsystem