Remote Control Commands
R&S
®
SMBV100A
617Operating Manual 1176.8016.02 ─ 17
Parameters:
<Offset> float
Range: -1 to 1
Increment: 0.001
*RST: 0
Default unit: FS
Example:
BB:ARB:TSIG:RECT:OFFS 0.5
sets the DC component
Manual operation: See "Offset DC (Rect Test Signal)" on page 406
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles <Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
<Samples> integer
Range: 4 to 1000
*RST: 100
Example:
BB:ARB:TSIG:RECT:SAMP 400
sets 400 samples per period
Manual operation: See "Samples per Period (Rect Test Signal)" on page 405
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:FREQuency <Frequency>
Sets the frequency of the simple sinusoidal test signal. This signal is used as output via
the I channel. A sine wave of the same frequency but optionally phase-shifted is gener-
ated on the Q path (ARB:TSIGnal:SINE:POFFset).
Parameters:
<Frequency> float
Range: 100 to depends on hardware
Increment: 0.01
*RST: 1000
Default unit: Hz
Example:
BB:ARB:TSIG:SINE:FREQ 100 kHz
sets a sine signal of 100 kHz.
Manual operation: See "Frequency (Sine Test Signal)" on page 404
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe <Phase>
The command sets the phase offset of the sine wave on the Q channel relative to the
sine wave on the I channel.