Remote Control Commands
R&S
®
SMBV100A
623Operating Manual 1176.8016.02 ─ 17
Parameters:
<Reference> RMS | PEAK
*RST: RMS
Manual operation: See "Power Reference" on page 428
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:STATe <State>
Enables/disables the selected carrier.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOURce1:BB:ARBitrary:MCARrier:CARRier15:STATe
ON
Manual operation: See " State " on page 435
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CFACtor:MODE <Mode>
The command sets the mode for optimizing the crest factor by calculating the carrier
phases.
Parameters:
<Mode> OFF | MIN | MAX
OFF
There is no automatic setting for minimizing or maximizing the
crest factor. The Phase setting (command
BB:ARB:MCAR:CARR:PHAS) is in use.
MIN
The crest factor is minimized by internally calculating optimized
carrier phases. The Phase setting (command
BB:ARB:MCAR:CARR:PHAS) is invalid.
MAX
The crest factor is maximized by internally calculating optimized
carrier phases. The Phase setting (command
BB:ARB:MCAR:CARR:PHAS) is invalid.
*RST:
OFF
Example:
BB:ARB:MCAR:CFAC:MODE OFF
switches off automatic crest factor optimization. The setting
SOUR:BB:ARB:MCAR:CARR:PHAS has an effect.
Manual operation: See "Crest Factor Mode" on page 427
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CLIPping:CFACtor <CFactor>
Sets the value of the desired crest factor, if baseband clipping is enabled
(BB:ARB:MCAR:CLIP:STAT ON).