Remote Control Commands
R&S
®
SMBV100A
655Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:RDELay?..................................................... 660
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit.................................................. 660
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence.................................................................. 661
[:SOURce<hw>]:BB:DM:TRIGger:ARM:EXECute
The command stops signal generation for trigger modes Armed_Auto and Armed_Ret-
rigger. A subsequent internal or external trigger event restart signal generation.
Example:
BB:DM:TRIG:SOUR INT
sets internal triggering.
BB:DM:TRIG:SEQ ARET
sets Armed_Retrigger mode, i.e. every trigger event causes sig-
nal generation to restart.
BB:DM:TRIG:EXEC
executes a trigger, signal generation is started.
BB:DM:TRIG:ARM:EXEC
signal generation is stopped.
BB:DM:TRIG:EXEC
executes a trigger, signal generation is started again.
Usage: Event
Manual operation: See "Arm" on page 373
[:SOURce<hw>]:BB:DM:TRIGger:EXECute
Executes a trigger. The internal trigger source must be selected using the com-
mand:BB:DM:TRIG:SOUR INT and a trigger mode other than AUTO must be
selected using the command :BB:DM:TRIG:SEQ.
Example:
BB:DM:TRIG:SOUR INT
sets internal triggering.
BB:DM:TRIG:SEQ RETR
sets Retrigger mode, i.e. every trigger event causes signal gen-
eration to restart.
BB:DM:TRIG:EXEC
executes a trigger.
Usage: Event
Manual operation: See "Execute Trigger" on page 362
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
(enabled for "Trigger Source" External)
Enables/disables output of the signal synchronous to the external trigger event.
For or two or more R&S SMBVs configured to work in a master-slave mode for syn-
chronous signal generation:
●
disable this parameter in the slave instruments, in case the master instrument pro-
vides the slaves with its internal trigger signal and
SOURce Subsystem