Remote Control Commands
R&S
®
SMBV100A
657Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay <Delay>
Defines the delay between the signal on the marker outputs and the start of the signal,
expressed in terms of symbols. Command [:SOURce<hw>]:BB:DM:TRIGger:
OUTPut:DELay:FIXed can be used to restrict the range of values to the dynamic
range, i.e. the range within which a delay of the marker signals can be set without
restarting the marker and signal.
Parameters:
<Delay> float
Range: 0 to 16777215
Increment: 0.001
*RST: 0
Example:
BB:DM:TRIG:OUTP2:DEL 16
sets a delay of 16 symbols for the signal on connector MARKER
2.
Manual operation: See "Marker x Delay" on page 376
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay:MINimum?
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay:MAXimum?
Queries the maximum marker delay for setting :BB:DM:TRIG:OUTP:DEL:FIX ON.
Return values:
<Maximum> float
Range: 0 to 16777215
Increment: 1E-3
*RST: 2000
Example:
BB:DM:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
BB:DM:TRIG:OUTP:DEL:MAX
queries the maximum of the dynamic range.
Response: 2000
the maximum for the marker delay setting is 2000 symbols.
Usage: Query only
Manual operation: See "Marker x Delay" on page 376
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
Queries the status of signal generation for all trigger modes with digital modulation on.
SOURce Subsystem