Remote Control Commands
R&S
®
SMBV100A
659Operating Manual 1176.8016.02 ─ 17
Parameters:
<Source> INTernal|EXTernal
INTernal
manual trigger or *TRG.
EXTernal
trigger signal on the TRIG connector.
*RST: INTernal
Example:
SOURce1:BB:DM:TRIGger:SOURce EXTernal
sets external triggering via the TRIGGER connector.
Manual operation: See "Trigger Source" on page 374
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT <TrigDelUnit>
Determines the units in that the trigger delay is expressed.
Parameters:
<TrigDelUnit> SAMPle | TIME
*RST: SAMPle
Example:
SOURce1:BB:DM:TRIGger:SEQuence RETR
SOURce1:BB:DM:TRIGger:SOURce EXT
SOURce1:BB:DM:TRIGger:DELay:UNIT TIME
SOURce1:BB:DM:TRIGger:EXTernal:TDELay 100
SOURce1:BB:DM:TRIGger:EXTernal:RDELay?
// Response: 100
SOURce1:BB:DM:TRIGger:DELay:UNIT SAMPle
SOURce1:BB:DM:TRIGger:EXTernal:DELay?
// Response: 27.08
Manual operation: See "External Delay Unit" on page 375
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:DELay <Delay>
Sets the trigger delay on external triggering.
Parameters:
<Delay> float
Range: 0 to 65535
Increment: 0.01
*RST: 0
Default unit: symbol
Example:
BB:DM:TRIG:SOUR EXT
selects an external trigger
BB:DM:TRIG:DEL 50
sets a delay of 50 symbols for the trigger.
Manual operation: See "Trigger Delay" on page 375
SOURce Subsystem