Remote Control Commands
R&S
®
SMBV100A
683Operating Manual 1176.8016.02 ─ 17
Manual operation: See "Load User Mapping" on page 357
[:SOURce<hw>]:BB:DM:MLISt:POINts?
Queries the user modulation mapping list length.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example:
BB:DM:FORM USER
selects the User modulation type.
BB:DM:MLIS:POIN?
queries the list length.
Usage: Query only
[:SOURce<hw>]:BB:DM:MLISt:SELect <Filename>
Selects the user mapping list in the default directory. The default directory is set using
command MMEM:CDIRectory. A path can also be specified, in which case the files in
the specified directory are selected. If a user mapping list with the specified name does
not yet exist, it is created. The file extension may be omitted. Only files with the file
extension *.vam will be selected or created.
Parameters:
<Filename> string
Example:
BB:DM:MLIS:SEL 'c_list1'
selects the user mapping list c_list1.
Manual operation: See "Load User Mapping" on page 357
[:SOURce<hw>]:BB:DM:SETTing:CATalog?
Reads out the files with Custom Digital Modulation settings in the default directory. The
default directory is set using command MMEM:CDIRectory. Only files with the file
extension *.dm will be listed.
Return values:
<Catalog> string
Example:
MMEM:CDIR '/var/user/dig_mod'
sets the default directory.
BB:DM:SETT:CAT?
reads out all the files with Custom Digital Modulation settings in
the default directory.
Response: 'DM_1'
the file DM_1 with Custom Digital Modulation settings is availa-
ble.