Remote Control Commands
R&S
®
SMBV100A
685Operating Manual 1176.8016.02 ─ 17
Manual operation: See "Save/Recall" on page 353
[:SOURce<hw>]:BB:DM:SETTing:STORe:FAST <Fast>
Determines whether the instrument performs an absolute or a differential storing of the
settings.
Enable this function to accelerate the saving process by saving only the settings with
values different to the default ones.
Note: This function is not affected by the "Preset" function.
Parameters:
<Fast> 0 | 1 | OFF | ON
*RST: 1
Manual operation: See "Fast Save" on page 162
[:SOURce<hw>]:BB:DM:STANdard:ULISt:CATalog?
Reads out the files with Digital Standard settings in the default directory. The default
directory is set using command MMEM:CDIRectory. Only files with the file extension
*.dm_stu will be listed.
Return values:
<Catalog> string
Example:
MMEM:CDIR '/var/user/dig_mod'
sets the default directory.
BB:DM:STAN:ULIS:CAT?
reads out all the files with Digital Standard settings in the default
directory.
Response: 'DM_QAM'
the file DM_QAM with Digital Standard settings is available.
Usage: Query only
Manual operation: See "Save/Recall User" on page 356
[:SOURce<hw>]:BB:DM:STANdard:ULISt:DELete <Filename>
Deletes the selected file with Digital Standard settings. The directory is set using com-
mand MMEM:CDIRectory. A path can also be specified, in which case the files in the
specified directory are read. The file extension may be omitted. Only files with the file
extension *.dm_stu will be deleted.
Setting parameters:
<Filename> string
Example:
BB:DM:STAN:ULIS:DEL 'DM_QAM'
deletes file 'DM_QAM'.
Usage: Setting only