Remote Control Commands
R&S
®
SMBV100A
751Operating Manual 1176.8016.02 ─ 17
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
IQ:OUTP:ANAL:BIAS:OPT:STAT ON
activates optimization for RF output.
Manual operation: See "Optimize I/Q Signals for RF Output" on page 317
[:SOURce<hw>]:IQ:OUTPut:LEVel <Level>
Sets the off-load voltage.
Single ended Output:
●
Value range: 1 mV ... 1.5 V EMF.
Differential Output:
●
Value range: 2 mV ... 3 V EMF.
Parameters:
<Level> float
Range: 0.02 to 4
Increment: 0.001
*RST: 1
Default unit: V
Example:
SOURce1:IQ:OUTPut:LEVel 2
sets a off-load level of 2 V
Manual operation: See "I/Q Level Vp (EMF)" on page 318
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP:RESult?
Displays the resulting output off-load voltage of both signal components at the rear
panel.
Return values:
<Result> float
Range: 0.02 to 4
Increment: 0.001
*RST: 1
Usage: Query only
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE <Mode>
Selects the mode for setting the outputs.
Selection FIXed locks all IQ:OUTPut-commands except for IQ:OUTPut:TYPE and
sets them to predefined values ( *RST).
SOURce Subsystem