Remote Control Commands
R&S
®
SMBV100A
752Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode> FIXed | VARiable | VATTenuated
FIXed
The settings are fixed and cannot be changed.
VARiable
The settings can be changed (command [:SOURce<hw>]:IQ:
OUTPut:LEVel on page 751).
VATTenuated
Enables a precise digital attenuation of the signal. The level of
the baseband and the RF signal is attenuated by the value set
with the command [:SOURce<hw>]:IQ:OUTPut[:ANALog]:
POWer:DATTenuation.
*RST:
FIXed
Manual operation: See "Mode" on page 317
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:DATTenuation
<DAttenuation>
The level of the baseband and the RF signal is attenuated by this value.
Parameters:
<DAttenuation> float
Range: -3.522 to 80
Increment: 1E-3
*RST: 0
Example:
SOUR:IQ:OUTP:ANAL:MODE VATT
SOUR:IQ:OUTP:ANAL:POW:DATT 10dB
The level of the baseband and RF signal is attenuated with
10dB.
Manual operation: See "Digital Attenuation" on page 318
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE <Type>
Selects an input type.
Parameters:
<Type> SINGle | DIFFerential
SINGle
The I/Q-signal components are output single-ended.
DIFFerential
The I/Q-signal components are output differential.
*RST: SINGle
Example:
IQ:OUTP:ANAL:TYPE DIFF
the I/Q-signal components are output differential.
Manual operation: See "I/Q Output Type" on page 317
SOURce Subsystem