Remote Control Commands
R&S
®
SMBV100A
786Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:POWer:LMODe..................................................................................... 790
[:SOURce<hw>]:POWer:MANual.....................................................................................790
[:SOURce<hw>]:POWer:MODE.......................................................................................791
[:SOURce<hw>]:POWer:PEP?........................................................................................ 792
[:SOURce<hw>]:POWer:POWer......................................................................................792
[:SOURce<hw>]:POWer:SPC:CRANge............................................................................ 792
[:SOURce<hw>]:POWer:SPC:DELay............................................................................... 793
[:SOURce<hw>]:POWer:SPC:PEAK................................................................................ 793
[:SOURce<hw>]:POWer:SPC:SELect.............................................................................. 793
[:SOURce<hw>]:POWer:SPC:STATe...............................................................................794
[:SOURce<hw>]:POWer:SPC:TARGet............................................................................. 794
[:SOURce<hw>]:POWer:STARt.......................................................................................794
[:SOURce<hw>]:POWer:STEP[:INCRement].................................................................... 795
[:SOURce<hw>]:POWer:STEP:MODE............................................................................. 795
[:SOURce<hw>]:POWer:STOP........................................................................................796
[:SOURce<hw>]:POWer:ALC:OMODe <OffMode>
The command sets the level control mode which becomes active when automatic level
control is deactivated (ALC Off).
Parameters:
<OffMode> SHOLd | TABLe | HIGHaccuracy
SHOLd
Level control is activated briefly if the level or frequency changes
("ALC Off Sample & Hold").
HIGHaccuracy
The ALC mode High Accuracy ensures improved level accuracy.
Note: This mode increases the frequency and level setting time
up to 20 ms!
TABLe
Internal level control is performed according to the ALC table.
*RST: SHOLd
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:OMOD SHOL
level control is briefly activated if the frequency or level changes.
Manual operation: See " State - ALC " on page 227
[:SOURce<hw>]:POWer:ALC:SONCe
Temporarily activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage: Event
SOURce Subsystem