Remote Control Commands
R&S
®
SMBV100A
828Operating Manual 1176.8016.02 ─ 17
:SYSTem:COMMunicate:GPIB:LTERminator <LTerminator>
Sets the terminator recognition for remote control via GPIB bus.
Parameters:
<LTerminator> STANdard | EOI
EOI
The terminator must be sent together with the line message EOI
(End of Line). This setting is recommended for binary block
transmissions where a character could coincidentally have the
value LF (Line Feed) but is not intended as the terminator. This
setting must be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of
whether it is sent with or without EOI.
*RST: n.a. (factory preset: STANdard)
Example:
SYSTem:COMMunicate:GPIB:LTERminator EOI
Only a character which is sent simultaneously with the line mes-
sage EOI is accepted as the terminator.
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
Sets the GPIB address.
Parameters:
<Address> integer
Range: 0 to 30
*RST: n.a. (factory preset: 28)
Example:
SYSTem:COMMunicate:GPIB:SELF:ADDRess 28
Sets GPIB address.
Manual operation: See "GPIB channel address" on page 143
:SYSTem:COMMunicate:NETWork[:COMMon]:DOMain <Domain>
Sets the primary suffix, that is the DNS name without the host name part.
Parameters:
<Domain> string
Example:
SYSTem:COMMunicate:NETWork:COMMon:DOMain
'ABC.DE'
sets the domain of the network.
Manual operation: See "DNS Suffix" on page 139
:SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname <Hostname>
Sets the individual host name of the R&S SMBV.
SYSTem Subsystem