Remote Control Commands
R&S
®
SMBV100A
830Operating Manual 1176.8016.02 ─ 17
:SYSTem:COMMunicate:NETWork:IPADdress <IpAddress>
Sets the IP address.
Parameters:
<IpAddress> string
Range: 0.0.0.0. to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress '7.8.9.10'
sets the IP address of the instrument.
Manual operation: See "IP Address" on page 138
:SYSTem:COMMunicate:NETWork[:IPADdress]:GATeway <Gateway>
Sets the IP address of the default gateway.
Parameters:
<Gateway> string
Range: 0.0.0.0 to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
'1.2.3.4'
sets the IP address of the default gateway.
Manual operation: See "Default Gateway" on page 139
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK <Mask>
Sets the subnet mask.
Parameters:
<Mask> string
Example:
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:
MASK '255.255.0.0'
determines the subnet mask.
Manual operation: See "Subnet Mask" on page 138
:SYSTem:COMMunicate:NETWork:MACaddress <MacAddress>
Queries the MAC address of the network adapter.
Parameters:
<MacAddress> string
Example:
SYST:COMM:NETW:MAC
queries the MAC address.
Manual operation: See "MAC Address" on page 139