Remote Control Commands
R&S
®
SMBV100A
831Operating Manual 1176.8016.02 ─ 17
:SYSTem:COMMunicate:NETWork:STATus?
Queries the network configuration state.
Return values:
<State> 0 | 1 | OFF | ON
Usage: Query only
Manual operation: See "Network Status" on page 137
:SYSTem:COMMunicate:NETWork:RESTart
Restarts the network connection to the instrument, terminates the connection and sets
it up again.
Example:
SYSTem:COMMunicate:NETWork:RESTart
Usage: Event
Manual operation: See "Restart Network" on page 139
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command :SYSTem:COMMunicate:GPIB[:
SELF]:ADDRess.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:GPIB:RESource?
queries the VISA resource string.
Response: "GPIB::28::INSTR"'
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 144
:SYSTem:COMMunicate:NETWork:RESource?
Queries the VISA resource string, used for remote control of the instrument with
VXI-11 protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:NETWork:RESource?
Response: "TCPIP::192.1.2.3::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 144