Sirona Dental Systems GmbH 4 Operation
Operating Instructions SINIUS / SINIUS CS 4.5 Dentist element
båÖäáëÜ
Treatment, to rque adjustment
Endodontics, speed and torque settings
Contra-angle handpiece, endodontics
Turning on/off the automatic motor stop of the ApexLocator
Turning on/off t he automatic motor sto p of the ApexLocator
If your treatment center is equipped with the ApexLocator option, you can
determine that the motor stops at the physiological apex. The motor stop
can be combined with the
Auto-Reverse
function, see "Switching the
Auto-Reverse function on/off".
Touch the
Apex stop
key.
If the key is highlighted orange, the automatic motor stop is
activated.
Switching the auto reverse function on/off
Switching the auto reverse function on/off
A setting can be made so that the burr drive automatically switches to
counterclockwise rotation when the preset torque value is reached.
Auto-Reverse with ApexLocator
If your treatment center is equipped with the ApexLocator option, you can
determine that the burr drive stops automatically at the physiological
apex, see “Turning on/off the automatic motor stop of the
ApexLocator" [ 119]. If the Auto-Reverse function is switched on, the
next time the foot switch is activated following a motor stop, the motor is
switched to counterclockwise rotation. When the file is withdrawn, the
burr drive automatically switches back to clockwise rotation.
Touch the
AutoRev
key.
If the key is marked orange, the Auto Reverse function is
activated.
IMPORTANT
Torque adjustment
The maximum adjustable torque depends on the system motor and the
speed settings.
CAUTION
Improperly selected speeds and torque values endanger the patient.
Treatment errors, e.g. breaking of a file, may result from incorrect
settings.
Observe the manufacturer's instructions regarding file systems.
CAUTION
Only the Sirona ENDO 6:1 contra-angle handpiece may be used for the
endodontics function.
Instruments from other manufacturers can lead to malfunctions. Outside
instruments may be improperly calibrated for endodontics.
Use only the ENDO 6:1 contra-angle handpiece from Sirona for
endodontics.
Check whether the reduction ratio on the touchscreen agrees with
the value specified on the contra-angle handpiece being used.