4-20
IM DL950-02EN
Setting the State and Trigger Conditions (Pattern Setup)
2.
Tap Pattern Setup, or press the soft key. The following menu appears.
On a module with logic inputs
or sub channels, expand the
menu, and set each bit or sub
channel.
Set the state condition (H, L, or X (do not use as a trigger source)).
Set the trigger level.
Set the hysteresis ( , , ).
Set the determination mode.
Set the reference time.*
* Set T1 and T2 when the determination mode is T1 < T < T2 or T <T1, T2 < T.
Set Time when the determination mode is T < Time or T >Time.
Note
If linear scaling is applied to the trigger source, set the trigger level using a linearly scaled value.
Setting the Determination Mode (Mode)
Set what kind of relationship must be established between period T and the specified
reference times (Time or T1 and T2) for the instrument to trigger.
T < Time Period T must be shorter than the reference time (Time).
T > Time Period T must be longer than the reference time (Time).
T1 < T < T2 Period T must be longer than reference time T1 and shorter than reference time T2.
T < T1, T2 <T Period T must be shorter than reference time T1 or longer than reference time T2.
4.12 Triggering on a Period Trigger (Enhanced)