EasyManua.ls Logo

YOKOGAWA ScopeCorder DL950 - Triggering on A to B(N) (Enhanced)

YOKOGAWA ScopeCorder DL950
316 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-11
IM DL950-02EN
4.8 Triggering on an A to B(N) Trigger (Enhanced)
This section explains the following settings for triggering on an A to B(N) trigger:
Applicable in Scope Mode
Trigger source
State condition
State condition achievement condition
Number of times state condition B must be met
Trigger conditions (trigger level, hysteresis)
► “A to B(N) Trigger (Enhanced)” in the Features Guide
Trigger Menu
1.
On the waveform screen’s MENU, tap TRIGGER > MENU, or press TRIG.MENU. A trigger
menu appears.
Select Enhanced.
Set the trigger type to A to B (N).
See “Setting the State and
Trigger Conditions.”
Set the trigger position, trigger delay, and hold-off
► sec. 4.2
Set the trigger mode. ► sec. 4.1