EasyManua.ls Logo

YOKOGAWA ScopeCorder DL950 - Setting X-Y Waveform Trace

YOKOGAWA ScopeCorder DL950
316 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-2
IM DL950-02EN
6.1 Setting the X-Y Waveform Trace
This section explains the following settings for displaying X-Y waveforms:
Items Common to Scope Mode and Recorder Mode
Turning X-Y waveform display on and off, source waveform (X axis, Y axis)
Start and end points of display range
► “X-Y Waveform Setup (Setup)” in the Features Guide
X-Y menu
1.
On the waveform screen’s MENU, tap DISPLAY > X-Y, or press SHIFT+DISPLAY(X-Y). The
X-Y menu appears.
Turns the display on or off
Select the window that you want to edit.
See “Selecting Traces.”
Set the display range start and
end points.
Set the cursor. ► sec. 6.3
Set the pen marker. ► sec. 6.2
Displays the Next menu. ► sec. 6.2
Selecting Traces
2.
Tap Setup, or press the soft key. The following menu appears.
Turns the display on or off
For a 16-CH Temperature/Voltage Input
Module, set at the sub channel level.
Set the X-axis trace.
Set the Y-axis trace.
Set the serial bus input at the bit level.
Set XY1 to XY4 on Window1 and XY5 to XY8 on Window2.

Table of Contents