5-5
IM DL950-02EN
Interpolation method
You can select the interpolation method from the following:
OFF(2k): Interpolation is not performed. If outside the interpolation area and the display record
length exceeds 2 kpoint, the data is decimated down to 2 kpoint and displayed. If the
display record length is less than or equal to 2 kpoint, all the points are displayed.
OFF(100k): Interpolation is not performed. If outside the interpolation area and 100 kpoint is
exceeded, the data is decimated down to 100 kpoint and displayed. If the display
record length is less than or equal to 100 kpoint, all the points are displayed.
Sine: Sampling data is interpolated with a sine curve.
Line: Sampling data is interpolated linearly.
Pulse: Sampling data is interpolated stepwise.
5.3 Setting the Grid and Interpolation Method