10-4
IM DL950-02EN
Setting Measurement Items (Item Setup)
4.
Tap Item Setup, or press the soft key. The following screen appears.
Select the measurement items that you want to use.
X1: Time value at Cursor1
X2: Time value at Cursor2
ΔX: Difference between the time values of Cursor1 and Cursor2
1/ΔX: Reciprocal of the difference between the time values of Cursor1
and Cursor2
Y1: Vertical value at the intersection of Cursor1 and the waveform
Y2: Vertical value at the intersection of Cursor2 and the waveform
ΔY: Difference in the vertical values at the intersection of
Cursor1/Cursor2 and the waveform
Cursor Value Read Mode
4.
Tap Other Setup, or press the soft key.
5.
Tap Cursor Read Mode, or press the soft key to set the cursor value read mode.
The two modes available for reading Y1 and Y2 measurements are Display (display data) or
ACQ (acquisition data).
Display: The screen is divided into 1001 parts, and the maximum and minimum values
displayed on those rasters are read.
ACQ: Measurement data acquired at the sample rate of each waveform is read. When all
traces are selected, measurement data acquired at the sample rate calculated from
T/div and record length is read.
10.2 Measuring with Vertical Cursors

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ScopeCorder DL950 and is the answer not in the manual?

YOKOGAWA ScopeCorder DL950 Specifications

General IconGeneral
BrandYOKOGAWA
ModelScopeCorder DL950
CategoryTest Equipment
LanguageEnglish

Related product manuals