15-3
IM DL950-02EN
Setting the Action to Execute (Action Setup)
2.
Tap Action Setup, or press the soft key. The following screen appears.
Saves the waveform data
Set the waveform parameter save
destination.
Saves a screen capture
Saves waveform parameters measured
automatically
Set mail transmission and the number
of mail transmissions.
Set the beep sound.
Screen capture data format
(PNG, BMP, JPEG)
Set the auto file naming method
(Date, Numbering).
Set the waveform parameter file name.
Waveform data format
Set the creation of date file folders.
When Dual Capture Mode Is Off
When Dual Capture Mode Is On
Low-speed sampling waveform data format
High-speed sampling waveform data format
Note
When SSD recording is enabled, you cannot set actions.
Executing Actions
After you set the action mode, the actions to execute, and the number of times to execute the
actions, press START/STOP. The actions will be executed each time that the Instrument triggers.
“ACTION” appears at the top center of the screen when an action is being executed. To stop, press
START/STOP.
Note
• If the action to execute is mail transmission and the specified number of mail transmissions is less than
the number of specified actions, mail transmission will stop after the specified number of transmissions.
If the specified number of mail transmissions is greater than the number of specified actions, mail
transmission will stop after the specified number of actions.
• The action settings are shared with the action settings of GO/NO-GO.
15.1 Setting Actions