21-5
IM DL950-02EN
Configuring IEEE1588 Settings (IEEE1588)
4.
Tap IEEE1588, or press the soft key. The following screen appears.
Set the time synchronization.
► sec. 21.1
Select Master/Slave.
Set the master clock source.
Set Priority1 (0 to 255).
Set Priority2 (0 to 255).
Set the leap second (0 to 999).
IEEE1588 information display
Master Source GPS is available on models with the IRIG/GPS option (/C35).
In IEEE1588, the master clock in the network is selected according to the rules stipulated in the
IEEE1588 standard, taking into account the Piority1 and Priority2 values.
Devices with smaller Priority1 and Priority2 values have priority in selecting the master clock.
21.2 Configuring Master Settings (/C40 option)