4-4
IM DL950-02EN
4.2 Setting the Trigger Position, Trigger Delay, and
Hold-off
This section explains the following trigger settings:
Applicable in Scope Mode
Trigger position
Trigger delay
Hold-off
► “Trigger Position (Position),” “Trigger Delay (Delay Time),” and
“Trigger Hold-off (Hold Off)” in the Features Guide
POSITION/DELAY Menu
1.
On the waveform screen’s MENU, tap TRIGGER > MENU, or press TRIG.MENU. A trigger
menu appears.
2.
Tap Position/Delay, or press the soft key. The following menu appears.
Set the position and
delay.
Trigger menu
Set the trigger position to
a position of your choice.
Set the trigger position
to 10%.
Set the trigger position
to 50%.
Set the trigger position
to 90%.
See “Setting the Hold-off
Time.”
See “Setting the Trigger
Delay.”
Set the trigger mode.
► sec. 4.1
Select Simple or Enhanced.
Set the trigger source.
► sec. 4.3
Set the trigger slope.
► sec. 4.3
Set the trigger level.
► sec. 4.3
Set the trigger hysteresis.
► sec. 4.3
Setting the Hold-off Time (Hold Off)
The trigger hold-off feature temporarily stops the detection of the next trigger once a trigger has
occurred.
Setting the Trigger Delay (Delay Time)
The trigger delay is the delay from the trigger point until waveforms are displayed.
Note
You cannot specify a trigger delay when an external clock is being used as the time base.