EasyManua.ls Logo

Agilent Technologies Nano Indenter G200 - Tools Menu; Calibrate

Agilent Technologies Nano Indenter G200
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NanoSuite 4
Agilent Nano Indenter G200 User’s Guide 4-26
Tools Menu
The Tools menu, shown in Figure 4-28, is used to perform the
microscope-to-indenter calibration, validate methods, and recalculate
samples/tests.
Figure 4-28Tools menu in NanoSuite
The current test is the test that has its number highlighted in the Legend
Panel, such as the highlighted test shown in Figure 4-29, indicating test
number 10 in this example.
Figure 4-29List of current tests in Legend panel
Calibrate…
This command should only be used if you have purchased a calibration
kit and are familiar with its use or when required during diagnostics and
troubleshooting with a certified Agilent technician. This command may
be grayed out if the calibration kit is not installed.

Table of Contents

Related product manuals