Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide 6-31
Frame Stiffness and Area Function Calibration
The first step in calibrating the frame stiffness and area function is to
acquire data on fused silica over a range of displacements.
There are three methods from which the user may determine machine
compliance. These methods are:
• Frame Stiffness and Area Function Calibration: Express Test using
rapid indentation.
• Frame Stiffness Calibration: ISO 14577 Method using quasi-static
indentation.
• Frame Stiffness Calibration: CSM Method using dynamic
indentation.
See Frame Stiffness and Indenter Area Function Theory in Chapter 7,
"Theory" for more theoretical information.
General Information
Frame stiffness and area function information are stored in the same
location. To access the information, choose Tip > Configure Tip from
the main menu as shown in
Figure 6-45.
Figure 6-45Select Configure Tip from the Tip menu
Tip definitions are stored by user-defined names. The recommended
convention to use is Serial No.
Date Method (ie TB12938 042812
Express Test) as shown in
Figure 6-46 on page 6-32.