Quick Start Guide 2
Agilent Nano Indenter G200 User’s Guide 2-11
Figure 2-10Select a method from the Open Method drop-down list
4 Click OK.
The method that is loaded will serve to illustrate the most basic
operations of the instrument. At their most basic level, Nano Indenters
impose a force on a surface and measure the resulting penetration. This
method contains the following instructions for each test:
• The indenter shall approach the surface at a rate of 25 nm/sec.
• Once the indenter is in contact with the surface, the Load On
Sample shall be increased to Maximum Load within the time
span Time to Load. Both Maximum Load and Time to Load may
be specified.
• At the Maximum Load, the Load On Sample shall be held
constant for 10
seconds.
• Finally, the indenter shall be withdrawn from the sample at arate
that is twice as fast as the loading rate.
Prepare a Sample for Testing
Before you test a sample, you must prepare it by mounting it on a
sample disk.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Nano Indenter G200 and is the answer not in the manual?

Agilent Technologies Nano Indenter G200 Specifications

General IconGeneral
BrandAgilent Technologies
ModelNano Indenter G200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals