Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide 6-45
Frame Stiffness and Area Function Calibration: Express Test
1 Load the fused silica standard into the NanoVision stage.
2 In NanoSuite, select Open Method… from the Method menu.
3 From the Filename drop-down box in Open Method dialog box,
select Express Test > Express Test Tip Calibration as shown in
Figure 6-64.
Figure 6-64Select Express Test method
4 Click OK.
5 Right-click anywhere in the handset area and select Nano HandSet
from the resulting pop-up menu, as shown in
Figure 6-65 on
page 6-46.