Instrument Reference 3
Agilent Nano Indenter G200 User’s Guide 3-14
Sample Height Adjustment
Adjust the sample height as described in "Adjust the Height of the
Samples" on page 2-19 to ensure that samples do not strike the indenter
when the sample tray is moved.
Mounting Specimens
Please see the Quick Start Guide to "Prepare a Sample for Testing" on
page 2-11. Also see Crystalbond Hot Melt.
Objectives
The objectives, Figure 3- 17, are used to change magnification of the
microscope. One of the standard microscope objectives provided with
the system is the 10
X objective. The 10 X objective is a DIN Achromat
standard objective for a 150
mm fixed-tube microscope. The 10 X
objective is provided as a low-magnification option for the G200 optics
system. The instrument also comes with a 40
X objective. Other
objectives are also available, including 20
X, 60 X, and 100 X. For more
information about these objectives, contact your Agilent service
representative.
Figure3- 1710 X and 40 X objectives included with the G200
Objective lenses for the Nano Indenter G200 optics system are mounted
on precision slides so that they can be interchanged easily. To insert an
objective, grasp the objective and slide it into the mounting rails on the
microscope until it snaps into place. Make sure that you insert the
objective into the microscope such that the word FRONT is facing you,
as shown in
Figure 3- 18 on page 3-15. To remove the objective, reverse
the procedure. Grasp the objective and slide it out of the mounting rails.