Theory 7
Agilent Nano Indenter G200 User’s Guide 7-27
If S
2
/P is normal but the modulus data are awry, as shown in Figure,
then the area function should be recalculated as described in
Frame
Stiffness and Area Function Calibration in Chapter 6, "Commonly Used
Procedures".
Figure 7-21Data indicate area function requires calibration