Dynamic Contact Module II B
Agilent Nano Indenter G200 User’s Guide B-7
9 Exit and restart NanoSuite.
10 Find the head-indicator icon on the Status Bar in NanoSuite.
11 Confirm the head-indicator icon is set to XP highlighted in red as
shown in
Figure B-11.
Figure B-11XP icon on Status Bar
12 Remove the locking pins. Test the pins to determine which one offers
the most resistance and remove that pin first.
13 Install the locking pins into their docking holes in the tip-change
tool.