Express Test I
Agilent Nano Indenter G200 User’s Guide H-3
Figure H-2Run button in Nanosuite software
6 Once testing is complete:
a Go to the main menu
b Select Tip and then
c Select Configure Tip
If prompted to add the current tip to your setup, click Yes .
7 Click Calibrate Tip and Frame.
The next prompt warns you that the sample will be modified and saved.
8 Click Yes.
9 Edit and/or verify the inputs on the right-hand side of the screen
(beneath the Calculate button) to be:
a DCM
b Express Test
c The nominal lead term supplied on the tip's certificate should be
used. If this certificate is unavailable, leave the input at its default
value of 24.5.
10 Click the Calculate button.
11 Click the Save button.
12 In the Configure Tip Parameters window, click OK.
Note more information about the meaning of these calibrations and
optional settings refer to
Chapter 6, "Commonly Used Procedures".