EasyManua.ls Logo

AIXTRON CRIUS II - Checking the EPV Valves; Checking the Pressure Sensor

Default Icon
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
crius_II_en_00, Edition 06/2010
170 CONFIDENTIAL AIXTRON - Dokumentation
Maintenance System Manual
Pneumatic system CRIUS II
6
6.12.3 Checking the pressure sensor
The pressure sensor causes an alarm if the working pressure decreases
below the set limit.
1 Check the pressure limit value at the pressure sensor
The pressure limit must be set to 4 bar
2 If necessary, adjust the pressure limit.
See OEM manual.
Fig. 6-32 Pressure sensor
A Pressure display
B Operating elements
6.12.4 Checking the EPV valves
1 Activate the EPV valve to be checked in AIXACT
The valves are labeled!
2 Check the activated EPV valve at the corresponding valve cluster
The LED at the activated valve must emit light
Fig. 6-33 Valve cluster
A EPV valves
B Pneumatic hoses

Table of Contents