EasyManua.ls Logo

AIXTRON CRIUS II - Checking the Water Flows

Default Icon
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
crius_II_en_00, Edition 06/2010
180 CONFIDENTIAL AIXTRON - Dokumentation
Maintenance System Manual
Cooling water system CRIUS II
6
Removing the filter 6 If the back washing procedure does not clean the filter properly, dis-
mount the filter
7 Thoroughly clean the filter with a brush and water or replace it if neces-
sary
6.14.3 Checking the water flows
The flow values can be checked at the Data Sampler.
1 Open the reactor cabinet to access the Data Sampler
Fig. 6-40 Data Sampler
Checking the Data
Sampler configu-
ration
2 Check the configuration of the Data Sampler
All required channels must be activated
Checking the flows 3 Read the flow values in AIXACT or at the Data Sampler
4 Compare the actual flow values with the set values
5 If necessary, correct the flow by means of the corresponding manual
valve
The valves are labeled
NOTE:
For assignment of the flows refer to the provided water schematic.

Table of Contents