Chapter 2 SCPI Device Message Details
2-94
[:SENSe]:BANDwidth|:BWIDth:VIDeo <freq>
Video Bandwidth
Function
This command sets the video bandwidth (VBW). When Couple Time/Freq.
Domain is set to Off, each parameter is specified in accordance with the
already-specified frequency axis or time axis measurement.
Command
[:SENSe]:BANDwidth|:BWIDth:VIDeo <freq>
Parameter
<freq>
Video bandwidth (VBW)
Range/Resolution 1 Hz to 10 MHz (1 to 3 sequences), 5 kHz, OFF
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default Value set in VBW Auto
Details
The automatic setting of the video bandwidth (VBW) is OFF, if the video
bandwidth is changed.
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
Example of Use
To set VBW to 3 kHz.
BAND:VID 3KHZ
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:BANDwidth:VIDeo
[:SENSe]:CHPower:BANDwidth:VIDeo
[:SENSe]:OBWidth:BANDwidth:VIDeo