Chapter 2 SCPI Device Message Details
2-362
:CALCulate:ACPower:CORRection:NOISe[:AUTO] ON|OFF|1|0
Adjacent Channel Power Noise Cancel
Function
This command sets whether to enable the noise canceling function.
Refer to
[:SENSe]:ACPower:CORRection:NOISe[:AUTO].
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]?
Adjacent Channel Power Noise Cancel Query
Function
This command queries whether the noise canceling function is enabled.
Query
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]?
Response
<switch>
Parameter
<switch>
Type of result display
1
Enables the noise canceling function.
0
Disables the noise canceling function.
Details
This command is fixed to Off and cannot be set in the following cases:
•
ACP is Off.
•
Standard is Off.
•
Standard Parameter which can execute the noise canceling function is
not set in Load Standard Parameter.
•
Any of Span, RBW, Detection, Sweep Time, VBW (when Detection is
not set to RMS), and VBW Mode (when VBW is not set to Off and when
Detection is not set to RMS) has been changed from Standard
Parameter.
•
Scale Mode is Linear.