2.7 Sweep/Trigger/Gate
2-275
2
SCPI Device Message Details
:TRIGger[:SEQuence]:VIDeo:LEVel[:LOGarithmic]?
Log Scale Video Trigger Level Query
Function
This command queries the threshold of the level where trigger sweep
starts in video trigger in Log scale mode.
Query
:TRIGger[:SEQuence]:VIDeo:LEVel[:LOGarithmic]?
Response
<level>
Parameter
<level>
Threshold of the level where trigger sweep starts
Range –150 to +50 dBm
Resolution 1 dB
Suffix code None. Value is returned in dBm units.
Example of Use
To query the threshold of the level where trigger sweep starts.
TRIG:VID:LEV?
> –10
:TRIGger[:SEQuence]:VIDeo:LEVel:LINear <level>
Linear Scale Video Trigger Level
Function
This command sets the threshold value of the level where trigger sweep
starts.
Command
:TRIGger[:SEQuence]:VIDeo:LEVel:LINear <level>
Parameter
<level>
Threshold value of level where trigger
sweep starts
Range 0 to 100%
Resolution 1%
Suffix code None
Default 60%
Example of Use
To set the threshold of the level where trigger sweep starts to 50%.
TRIG:VID:LEV:LIN 50