Chapter 2 SCPI Device Message Details
2-68
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINear:PDIVision <percent>
Linear Scale Range
Function
This command sets the Y-axis scale magnification for the linear scale
display.
Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINear:PDIVision
<percent>
Parameter
<percent>
Y-axis scale magnification
1
1 %/div
2
2 %/div
5
5 %/div
10
10 %/div
Default value 10 %/div
Example of Use
To set the scale magnification to 5%/div.
DISP:WIND:TRAC:Y:LIN:PDIV 5
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINear:PDIVision?
Linear Scale Range Query
Function
This command queries the Y-axis scale magnification when Scale Mode is
set to Linear.
Query
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINear:PDIVision?
Response
<percent>
Parameter
<percent>
Y-axis scale magnification
1
1 %/div
2
2 %/div
5
5 %/div
10
10 %/div