2.15 Setting the TOI Measurement Functions
2-845
2
SCPI Device Message Details
:FETCh:TOI[n]?
TOI Read Fetch
Function
This command outputs the measurement result of the TOI measurement.
Query
:FETCh:TOI[n]?
Response
<tracedata_1>,<tracedata_2>,...
(
n=0
)
<worst_toi>,<worst_toi_freq>,
<lower_toi>,<lower_3rd_freq>,
<upper_toi>,<upper_3rd_freq>,
(
n=1
)
<worst_toi_freq>,<worst_3rd_abs>,<worst_toi>,
<lower_tone_freq>,<lower_tone_abs>,
<upper_tone_freq>,<upper_tone_abs>,
<lower_3rd_freq>,<lower_3rd_abs>,<lower_toi>,
<upper_3rd_freq>,<upper_3rd_abs>,<upper_toi>,
<worst_3rd_rel>,<lower_3rd_rel>,<upper_3rd_rel>
(n=2)
Parameter
<worst_toi>
Worst value of TOI
<worst_toi_freq>
3rd tone frequency that became the worst value
of TOI
<worst_3rd_abs>
Worst value of 3rd tone level
<lower_toi>
TOI of the lower frequency
<lower_3rd_freq>
Lower 3rd tone frequency
<lower_tone_freq>
Lower tone frequency
<lower_tone_abs>
Lower tone absolute level
<lower_3rd_abs>
Lower 3rd tone absolute level
<upper_toi>
TOI of the upper frequency
<upper_3rd_freq>
Upper 3rd tone frequency
<upper_tone_freq>
Upper tone frequency
<upper_tone_abs>
Upper tone absolute level
<upper_3rd_abs>
Upper 3rd tone absolute level
<worst_3rd_rel>
Worst value of 3rd tone relative level
<lower_3rd_abs>
Lower 3rd tone relative level
<upper_3rd_abs>
Upper 3rd tone relative level