Chapter 2 SCPI Device Message Details
2-604
[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1][:RESolution]:AUTO
<switch_1>[,<switch_2>[,<switch_3>[,<switch_4>[,<switch_5>[,<switch_6>[,<
switch_7>[,<switch_8>[,<switch_9>[,<switch_10>[,<switch_11>[,<switch_12>]
]]]]]]]]]]
Spectrum Emission Mask Offset Resolution Bandwidth Auto/Manual
Function
This command enables/disables the automatic resolution bandwidth
(RBW) setting function for the offset during Spectrum Emission Mask
measurement.
Command
[:SENSe]:SEMask:OFFSet[1]:LIST:BANDwidth[1][:RESolution]
:AUTO <switch_n>
Parameter
<switch_n>
Automatic resolution bandwidth setting function
On/Off when measuring Offset-n
ON|1
Enables the automatic resolution bandwidth
setting function.
OFF|0
Disables the automatic resolution bandwidth
setting function.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To enable the automatic resolution bandwidth setting function for the
offset.
SEM:OFFS:LIST:BAND:AUTO
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON