2.10 Burst Average Power
2-453
2
SCPI Device Message Details
[:SENSe]:BPOWer|:TXPower:BURSt:STOP <time>
Burst Average Power Stop Time
Function
This command specifies the stop position (time) for Burst Average Power
measurement.
Command
[:SENSe]:BPOWer|:TXPower:BURSt:STOP <time>
Parameter
<time>
Start position (time)
Range Within the frequency range of trace display
Resolution
Formula for calculating the resolution is:
Sweep Time / (Trace Point – 1)
(Minimum value: 1ns)
Suffix code NS,US,MS,S
S is used when omitted.
Default 100 ms
Example of Use
To set the stop position (time) for Burst Average Power measurement to
200 ms.
BPOW:BURS:STOP 200MS
Related command
This command has the same function as the following command.
:CALCulate:BPOWer|:TXPower:BURSt:STOP
:CALCulate:BPOWer|:TXPower:BURSt:STOP <time>
Burst Average Power Stop Time
Function
This command specifies the stop position (time) for Burst Average Power
measurement. Refer to
[:SENSe]:BPOWer|:TXPower:BURSt:STOP.
Related command
This command has the same function as the following command.
[:SENSe]:BPOWer|:TXPower:BURSt:STOP