2.10 Burst Average Power
2-449
2
SCPI Device Message Details
:CALCulate:BPOWer|:TXPower:CORRection:NOISe[:AUTO] ON|OFF|1|0
Noise Cancel
Function
This command sets whether to enable the noise canceling function.
Refer to
[:SENSe]:BPOWer|:TXPower:CORRection:NOISe[:AUTO].
Related command
This command has the same function as the following command.
[:SENSe]:BPOWer|:TXPower:CORRection:NOISe[:AUTO]
[:SENSe]:BPOWer|:TXPower:CORRection:NOISe[:AUTO]?
Noise Cancel Query
Function
This command queries the On/Off state of the noise canceling function.
Query
[:SENSe]:BPOWer|:TXPower:CORRection:NOISe[:AUTO]?
Response
<switch>
Parameter
<switch>
Type of result display
1
Noise canceling function is enabled.
0
Noise canceling function is disabled.
Details
This command is not available in the following cases:
•
Burst Average Power is Off.
•
Standard is Off.
•
The Standard Parameter which can execute the noise canceling
function is not set in Load Standard Parameter.
•
Any of Span, RBW, Detection, Sweep Time, VBW (except the case that
Detection is RMS.), and VBW Mode (except the case that VBW is set to
Off and the case that Detection is RMS.) has been changed from
Standard Parameter.
•
Scale Mode is Linear.