2.13 Spectrum Emission Mask
2-623
2
SCPI Device Message Details
[:SENSe]:SEMask:SWEep[1]:TIME:AUTO?
Spectrum Emission Mask Sweep Time Auto/Manual Query
Function
This command queries the On/Off state of the automatic sweep time
setting function when measuring the reference power for Spectrum
Emission Mask measurement.
Query
[:SENSe]:SEMask:SWEep[1]:TIME:AUTO?
Response
<switch>
Parameter
<switch>>
Automatic sweep time setting function On/Off
state
1
Automatic sweep time setting function is
enabled.
0
Automatic sweep time setting function is
disabled.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To query the automatic sweep time setting function On/Off state when
measuring the reference power.
SEM:SWE:TIME:AUTO?
> 1