2.15 Setting the TOI Measurement Functions
2-847
2
SCPI Device Message Details
:MEASure:TOI[n]?
TOI Measure
Function
This command performs the TOI measurement and outputs the
measurement result.
This command has the same function as the following commands
executed in this order:
:CONFigure:TOI
:INITiate:TOI
:*WAI
:FETCh:TOI[n]?
:FETCh:TOI:IP3?
TOI Read Fetch IP3
Function
This command outputs the measurement result of the TOI measurement.
Query
:FETCh:TOI:IP3?
Response
<worst_toi>
Parameter
<worst_toi>
Worst value of TOI
No suffix code, in dB units, 0.01 dB resolution
(–999.0 is returned when there is an error and
no measurement is performed.)
Details
This command outputs the result of the last performed TOI
measurement. Because this function is not related to sweep, it can be
used to output the results of a previously finished measurement in a
different format.
To retry the measurement by using the results of another sweep attempt,
use the READ command.
Example of Use
To obtain the measurement result of the TOI measurement.
FETC:TOI:IP3?