2.7 Sweep/Trigger/Gate
2-251
2
SCPI Device Message Details
:INITiate:MODE:CONTinuous
Continuous Measurement
Function
This command sets the sweep mode to Continuous and starts continuous
sweep.
Command
:INITiate:MODE:CONTinuous
Example of Use
To start continuous sweep.
INIT:MODE:CONT
:INITiate:MODE:SINGle
Single Measurement
Function
This command sets the sweep mode to Single and starts the single
sweep.
Command
:INITiate:MODE:SINGle
Example of Use
To start single sweep.
INIT:MODE:SING
Details
To read out the measurement results after executing this command,
perform the synchronization control using the “*WAI” command.