Chapter 2 SCPI Device Message Details
2-314
[:SENSe]:SWEep:EGATe:VIEW:FREQuency <freq>
Gate View Frequency
Function
This command sets the center frequency for Gate View.
Command
[:SENSe]:SWEep:EGATe:VIEW:FREQuency <freq>
Parameter
<freq>
Center Frequency
Range
[MS269xA]
–100 MHz to 6.05 GHz (MS2690A)
–100 MHz to 13.6 GHz (MS2691A)
–100 MHz to 26.6 GHz (MS2692A)
[MS2830A]
–100 MHz to 3.7 GHz (Option 040)
–100 MHz to 6.1 GHz (Option 041)
–100 MHz to 13.6 GHz (Option 043)
–100 MHz to 26.6 GHz (Option 044)
–100 MHz to 43.1 GHz (Option 045)
[MS2840A]
–100 MHz to 3.7 GHz (Option 040)
–100 MHz to 6.1 GHz (Option 041)
–100 MHz to 26.6 GHz (Option 044)
–100 MHz to 44.6 GHz (Option 046)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default
[MS269xA]
3.00 GHz (MS2690A)
6.75 GHz (MS2691A)
13.25 GHz (MS2692A)
[MS2830A]
1.8 GHz (Option 040)
3.0 GHz (Option 041)
6.75 GHz (Option 043)
13.25 GHz (Option 044)
21.5 GHz (Option 045)
[MS2840A]
1.8 GHz (Option 040)
3.0 GHz (Option 041)
13.25 GHz (Option 044)
22.25 GHz (Option 046)
Details
This command is not available when Gate View is set to Off.
Example of Use
To set the center frequency to 123.456 kHz.
SWE:EGAT:VIEW:FREQ 123456