2.14 Spurious Emission Measurement Function
2-731
2
SCPI Device Message Details
[:SENSe]:SPURious:SEGMent:NUMBer <integer>
Edit Segment Number
Function
This command sets the number of the segment to set a parameter in.
Command
[:SENSe]:SPURious:SEGMent:NUMBer <integer>
Parameter
<integer>
Segment number
Range 1 to 20
Resolution 1
Example of Use
To set 3 for the segment to set a parameter in.
SPUR:SEGM:NUMB 3
[:SENSe]:SPURious:SEGMent:NUMBer?
Edit Segment Number Query
Function
This command queries the segment in which a parameter is set.
Query
[:SENSe]:SPURious:SEGMent:NUMBer?
Parameter
<integer>
Segment number
Range 1 to 20
Resolution 1
Example of Use
To query the segment in which a parameter is set.
SPUR:SEGM:NUMB?
> 3