Chapter
4
Native Device Message Details
4-144
Example of Use
To option the results of ACP measurement performed using the
parameter list file MyParam.xml.
MEAS:BATC:ACP? 1,"MyParam.xml"
>
0.0,-72.130,0.0,-72.130,-1.270,-73.400,-0.570,-72.700,-0
.780,-72.910,-1.030,-73.160,-999.0,-999.0,-999.0,-999.0
Format of Parameter list file
Refer to
:MEASure:ACP[n]?
in “Chapter 2 SCPI Device Message
Details”.
MEAS:BATC:IM?
Transmit Intermodulation Batch Measure
Function
This command executes Transmit intermodulation measurement and
outputs the result based on the specified parameter list file.
Query
:MEAS:BATC:IM? filename,spa_freq[,sg_freq[,device]]
Response
Two values of ref_carrier_a are output when ACP Reference is BSIDes.
ref_carrier_a(,ref_carrier_a),
lower_offset_1_rel,lower_offset_1_abs,
upper_offset_1_rel,upper_offset_1_abs,
lower_offset_2_rel,lower_offset_2_abs,
upper_offset_2_rel,upper_offset_2_abs,
lower_offset_3_rel,lower_offset_3_abs,
upper_offset_3_rel,upper_offset_3_abs,
spa_1_result,spa_2_result