2
SCPI Device Message Details
:CALCulate:LLINe[1]|2|3|4|5|6:ENVelope[:LEVel]:OFFSet <amplitude>
Envelope Offset
Function
This command sets the offset level (from the current peak Trace Point to
the Limit Point).
Command
:CALCulate:LLINe[1]|2|3|4|5|6:ENVelope[:LEVel]:OFFSet
<amplitude>
Parameter
<amplitude>
Offset level
Range –300.00 to 300.00 dB
Resolution 0.01 dB
Unit dB
Default 3.00 dB
Details
When the Create Envelope or Update Envelope function is executed, the
set value applies.
Example of Use
To set the offset level (from the peak Trace Point to the Limit Point) to 3
dB.
CALC:LLIN:ENV:OFFS 3